Start and Stop Recording Parameters

<< Click to Display Table of Contents >>

Navigation:  Software > Menu and Tool Commands > M.O.L.E. Menu > Setup Instrument >

Start and Stop Recording Parameters

When selecting the Advanced check box to display the Start and Stop Recording Parameters, refer to the descriptions below for detailed information.

mole_map-2810

1)If desired, configure a Start Parameter such as a threshold temperature or Delay Points by selecting the associated check box and entering the proper values. The maximum delay points in for a SuperM.O.L.E.® Gold 2, MEGAM.O.L.E.® 20 and V-M.O.L.E.® is 32767

Important Information

Specifying a threshold temperature “triggers” the recording process when any active channel reaches the specified temperature and Data Points “trigger” the M.O.L.E. Profiler to start recording when the specified data point is reached in the process. The actual delay is equal to the Interval times the Pts Dly.

2)If desired, configure a Stop Parameter by selecting the associated check box and entering the proper value. This parameter is a number of data points the M.O.L.E. Profiler will record during the data run. The maximum number is dependent on the number of channels turned on. The maximum points the SuperM.O.L.E.® Gold 2, MEGAM.O.L.E.® 20 and V-M.O.L.E.® can log at any recording interval is 9999 and the SuperM.O.L.E.® Gold is 5460.

Important Information

If a no stop parameter is specified, the M.O.L.E. Profiler will log continuously and will require the user to manually stop the recording process.