Charging the Power Pack

<< Click to Display Table of Contents >>

Navigation:  Sensor Platforms > MEGARIDER™ PTP® > Setup >

Charging the Power Pack

Because the M.O.L.E. Profiler is powered by a rechargeable Power Pack, it is important that it is charged and operating properly prior to performing every experiment. When using a MEGAM.O.L.E.® 20 or SuperM.O.L.E.® Gold a spare removable Power Pack may be ordered so one is charging while the other one is being used.

To charge the Power Pack:

1)Slide the interchangeable plug into the adaptor until it clicks indicating that is fully seated.
2)Plug the transformer end of the charger into a wall outlet and the USB connector end into the Data/Charging Port.

mm_charge

Symbol_(Info)

A completely discharged Power Pack takes about 8 hours to be fully charged. For quick charges, it can be charged for 15 minutes allowing one 10 minute data run to be performed.

3)When the charging cycle is complete, connect the Power Pack to the M.O.L.E. Profiler.

Important Information

The Power Pack can be charged continuously whenever the M.O.L.E. Profiler is not being used, however, if the M.O.L.E. Profiler is going to sit idle for five days or more, it should be removed from the charger.