Verify Process

<< Click to Display Table of Contents >>

Navigation:  Software > Menu and Tool Commands > Wizards Menu >

Verify Process

The Verify Process workflow is a wizard that configures a M.O.L.E. Profiler or OvenCHECKER™ to verify a process or machine performance using a Target 10 Specification file (*.T10).

Important Information

This is available in both Engineer & Verify Modes.

 

The Verify Process workflow:

1)Start the software by double clicking on the program icon.
2)Connect the M.O.L.E. Profiler to the computer. Refer to the Communications Setup topic for your specific M.O.L.E. Profiler for more information.
3)Set an Environment related for the type of experiment that is being performed. Either open an existing Environment Folder or create a new one.

Tip

If the desired Environment is not displayed on the Environment Sidebar, the user can use either the Open Environment Toolbar button or select the Open Environment Folder command from the File Menu.

mole_map-0025

4)On the File menu, click New. The Start dialog box appears with the five workflow wizard options.

mole_map-0190

5)On the Start dialog box, click the Verify Process command button and the workflow wizard appears.

Important Information

When navigating through the wizard, the step list on the left uses a color key to inform the user of the current step, steps that have been completed and remaining steps.

wizard_current

Current

wizard_complete

Completed

wizard_remain

Remaining

6)Select the desired instrument from the dialog box to choose the M.O.L.E. Profiler. If a M.O.L.E. Profiler has already been selected during a different process, the software automatically selects the M.O.L.E. Profiler connected to the COM Port previously used.

Important Information

If the software does not detect a M.O.L.E. Profiler, using the communication cable connect it to the computer and click the Scan for Instruments command button to search again. M.O.L.E.® MAP software allows multiple instruments to be connected to a computer at one time. Selecting the Scan for Instruments command button will detect all instruments and display them in the dialog box. If no instrument is detected the software displays all of the Demonstration thermal profilers to select from.

vmole-0040

7)Click the Next command button.
8)Select a Target 10 file from the list to verify. When placing the mouse cursor over a file on the list, a thumbnail image appears to help properly identify the assembly associated with it.

vmole-0050

Important Information

When selecting a Target 10 file to verify, the software displays files located in the \ECD\MegaMoleMAP\Target10Spec\ directory that are associated with the currently selected instrument.

9)Click the Next command button.
10)Verify the instrument status. This dialog box displays the health of the M.O.L.E. Profiler such as battery charge, internal temperature, thermocouple temperatures.

Important Information

If everything is OK, the dialog box displays a GREEN symbol. If there are any items that may prevent the user from collecting good data, they are highlighted and a yellow Warning symbol is displayed.

mole_map_status-go

mole_map_status-warning

OK

Warning

vmole-0060

11)Click the Next command button.
12)Review the machine settings and click the Next command button to continue.

mole_map-0200

13)Pass M.O.L.E. Profiler and test product through the process.

Important Information

If the selected M.O.L.E. Profiler is wireless RF, click the Next command button to proceed to the Step 14.

14)Press the OK button on the M.O.L.E. Profiler and wait for the GO or NO-GO indication (RED or GREEN). By pressing the OK button, the M.O.L.E. Profiler analyzes the most recent data run to verify the Target 10 Specifications.

smg2-ok

Important Information

Only the SuperM.O.L.E.® Gold 2, MEGAM.O.L.E.® 20 and V-M.O.L.E.® Profilers are equipped with the OK button. When using a M.O.L.E. Profiler other than those, please skip the previous step.

15)Connect the M.O.L.E. Profiler to the computer and click the Next command button.
16)Select the desired data run and then click the Next command button to read the data run from the M.O.L.E. Profiler.

Tip

This step of the wizard allows the user to remove a selected data run from the M.O.L.E. Profiler by either selecting the Delete After Reading check box or selecting the Delete command button and removing it before downloading. This feature is not available for the SuperM.O.L.E.® Gold Thermal Profiler as it does not have the ability to store more than one data run.

mole_map-0210

17)When the data run has been downloaded, the software prompts the user to save the verification data run file (*.XMG).

Important Information

If a data run (*.XMG) is saved in a different Environment Folder other than the currently selected, the software automatically activates the new Environment Folder. This process does not delete any data run files in the previously set Environment Folder and can be quickly accessed using the Recent Environment Folders on the File menu.

Save Data Run

 

18)When finished, click the Save command button.
19)Click the Next command button to show the Target 10-OK results.

mole_map-0220

20)Analyze the data by selecting the Target10-OK tab on the Profile Tab.

vmole-0070