OvenCHECKER

<< Click to Display Table of Contents >>

Navigation:  Software > Menu and Tool Commands > Wizards Menu > Create Target 10 Specification >

OvenCHECKER

The Create Target 10 Specification - OvenCHECKER is a wizard that helps the user create a Target 10 Specification using an OvenCHECKER™ data run that can be used to verify that your machine performance is the same in the future. To use this wizard, it requires the OvenCHECKER™ Oven Verification System which is powered by the V-M.O.L.E.® Profiler. Refer to topic OvenCHECKER™ for more information.

 

Create Target 10 Specification workflow:

Important Information

This is available when in Engineer Mode.

1)Connect the M.O.L.E. Profiler to the computer. Refer to the Communications Setup topic for your specific M.O.L.E. Profiler for more information.
2)Open the  OvenCHECKER™ barrier box and connect the M.O.L.E. Profiler to the Connector Bridge.
3)Set an Environment related for the type of experiment that is being performed. Either open an existing Environment Folder or create a new one.

Tip

If the desired Environment is not displayed on the Environment Sidebar, the user can use either the Open Environment Toolbar button or select the Open Environment Folder command from the File Menu.

ovenchecker_0005

4)Select the New (Start) button Start on the toolbar. A message box appears with the five workflow wizard options.

ovenchecker_0010

5)On the Start dialog box, select the Create T 10 using an OvenCHECKER radio button then click the Create T 10 Spec command button and the workflow wizard appears.
6)Select the instrument type that will be used to verify the process.

ovenchecker_0020

7)Confirm the assembly information such as part number, pallet size, sensor locations and a product image. This assembly information can then be saved from/to a an (*.XMA) file.

Save: This command button saves the current sensor locations and image to an  (*.XMA) file that is then available to be loaded for future data runs.

ovenchecker_0025

8)Click the Next command button.
9)Select your process specification. Select a Paste from the database or previously created Target 10 Specification. If your Paste does not appear in the database list click the New command button to create a new one. Refer to topic Create new Paste for more information.
10)Then choose the Profile Type (Ramp-Soak-Spike or Ramp-to-Spike). Ramp-Soak-Spike profile types are the only type allowed to be edited.

Tip

When the user selects a paste from the database, they can use the radio buttons below the drop down box to filter and display only the user created pastes from paste database.

ovenchecker_0030

Tip

The OvenCHECKER™ uses the begin temp, end temp and liquidous temp to measure the initial slope, soak, TAL (time above liquidous), peak parameters from the selected Paste or Target 10 Specification. Then based on the range of the parameters taken from the 3 OvenCHECKER™ sensors, automatically calculates the USL (Upper Specification Limits) and LSL (Lower Specification Limits).

11)Click the Next command button.
12)Select a machine from the machine drop down list. If your machine does not appear in the list, continue to set the machine information and select the Save As command button to save the new machine settings as an (*.XMR) file in the current Environment. Refer to topic Create new Machine for additional information.

mole_map-0027

This new machine file now appears in the machine drop down list for applying to new and existing data runs.

Tip

If the user has an existing machine (*.XMR) file that is not listed in the Machine drop down list, it can be loaded/imported by selecting the Load command button. Once loaded, this new machine file appears in the machine drop down list for applying to new and existing data runs.

ovenchecker_0035

13)Set the machine zone units.
14)Set the number of all machine zones (heating and cooling).
15)Set the machine temperature units.
16)Set the machine conveyor speed and units.
17)In the machine zone matrix, set the zone names (descriptions), zone lengths, zone temperatures (top / bottom) and zone types (heating or cooling).

ovenchecker_0040

18)Click the Next command button.
19)Verify the instrument status. This dialog box displays the health of the M.O.L.E. Profiler such as Power  Pack charge, internal temperature, thermocouple temperatures.

Important Information

If everything is OK, the dialog box displays a GREEN symbol. If there are any items that may prevent the user from collecting good data, they are highlighted and a yellow Warning symbol is displayed.

mole_map_status-go

mole_map_status-warning

OK

Warning

ovenchecker_0050

15)Review the machine settings and click the Next command button to continue.

ovenchecker_0060

16)After the oven stabilizes, turn the V-M.O.L.E.® Profiler on and press the record button.
17)Pass the OvenCHECKER™ through the machine.
18)Once the OvenCHECKER™ has been retrieved from the machine and the V-M.O.L.E.® Profiler has been stopped, disconnect the V-M.O.L.E.® Profiler from the OvenCHECKER™ and place it near the computer. Since you are creating a new Target 10 Specification, the OK button on the V-M.O.L.E.® does not need to be pressed at this time as contains an old Target 10 Specification.

Important Information

If the V-M.O.L.E.® Profiler already contains an previously recorded OvenCHECKER™ data run, a new data run does not need to be performed.

19)Restore the software and click the Next command button to read the data run from the M.O.L.E. Profiler.

Tip

This step of the wizard allows the user to remove a selected data run from the M.O.L.E. Profiler by either selecting the Delete After Reading check box or selecting the Delete command button and removing it before downloading. This feature is not available for the SuperM.O.L.E.® Gold Thermal Profiler as it does not have the ability to store more than one data run.

ovenchecker_0070

20)Once the OvenCHECKER™ data run is downloaded, review the calculated results of the Target 10 specification. The user can then decide if they wish to accept these or make final modifications to the Slope, Soak, Time Above and/or Peak parameter.

ovenchecker_0080

Important Information

If the user decides to make modifications, click the Modify Target 10 command button and the the Target 10-OK tab is displayed. Make the desired modifications to the Slope, Soak, Time Above and/or Peak parameter specifications. If any of the parameter specifications are changed the OK button will inform the user by displaying (OK Modified).

vmole-0020

21)Click the OK command button to accept the modified specifications or Cancel to discard them and return to the workflow wizard.

ovenchecker_0090

22)Click the Finish command button and the software prompts the user to save the Target 10 Specification file (*.T10).

Important Information

If a data run (*.XMG) is saved in a different Environment Folder other than the currently selected, the software automatically activates the new Environment Folder. This process does not delete any data run files in the previously set Environment Folder and can be quickly accessed using the Recent Environment Folders on the File menu.

ovenchecker_0100

23)When finished naming the file, click the Save command button to complete the workflow wizard. This step saves both the Target 10 Specification file (*.T10) and the downloaded data run which is displayed on the Target10-OK tab.

 

This OvenCHECKER™ Target 10 Specification can now be used to verify that your machine performance is the same in the future. Refer to topic Verify Process for more information.