Specifications

<< Click to Display Table of Contents >>

Navigation:  Instruments > SuperM.O.L.E.® Gold >

Specifications

SuperM.O.L.E.® Gold Profiler:

INPUTS:

Up to 6 type K ECD Micro-Thermocouples

PHYSICAL DIMENSIONS:

3.5" x 6" x 0.37"

TEMPERATURE MEASUREMENT RANGE:

-129°C to +1250°C (200°F to +2282°F)

INTERNAL OPERATING TEMPERATURE RANGE:

0°C to 65°C (32°F to 150°F)

Absolute Maximum Warranteed Internal Temperature:

82°C (180°F)* SuperM.O.L.E.® Gold automatically stops monitoring when the internal temperature exceeds 80°C (176°F)

Warning

*WARNING: Exceeding this temperature may permanently damage the equipment!

ACCURACY:

Within ± 0.1%+1°C. at –73, 100, 250 and 1000°C. Channel to channel compliance within 1°C.

PROGRAMMABLE SAMPLING PERIOD:

0.1 seconds to 24 hours

HARDWARE REAL-TIME CLOCK/CALENDER:

±1 minute/month at 25°C

OPEN THERMOCOUPLE DETECTION.

All six channels

NUMBER OF SAMPLES PER CHANNEL

5460 total for six channels

 

Power Pack:

POWER PACK CHARGE:

Typical 50 ten-minute profiles.

CHARGE TIME:

14 hours.

EXPECTED POWER PACK LIFE:

300-400 charging cycles.

OPERATING RANGE:

5.1V to 4.9V (the unit may become unreliable below 4.5V).

 

Power Pack Charger Model Specifications:

NORTH AMERICA:

ECD Part No. E31-0900-25

Input: 120 V, AC, 50/60 Hz, 2.8 VA Output: 9V DC, 120-200mA, 1.08VA

CONTINENTAL EUROPE:

ECD Part No. E31-0900-21

Input: 230 V, AC, 50/60 Hz, 2.8 VA Output: 9V DC, 120-200mA, 1.08VA

 

Thermocouple Specifications:

Thermocouples: Type K, Micro-Connector, Glass insulated. (Other insulation types available)

TIME CONSTANT*

Wire Size

mm (inches)

Still Air

427°C/38°C

800°F/100°F

60 Ft./Sec. Air

427°C/38°C

800/100°F

Still H2O

93°C/38°C

200°F/100°F

0.025 (0.001)

0.125 (0.005)

0.381 (0.015)

0.05 sec.

1.0 sec.

10.0 sec.

0.004 sec.

0.08 sec.

0.80 sec.

0.002 sec.

0.04 sec.

0.40 sec.

* The time constant is defined as the time required to reach 63.2% of an instantaneous temperature change.

 

Environmental Limitation Specifications:

Altitude up to 2000 meters

Maximum relative humidity 80% RH from 0°C to 31°C, decreasing linearly to 17% RH at 50°C.

Pollution Degree 2 (Normally only dry pollution, but with temporary conductivity caused by condensation)

Power Pack Charger Only:

Mains Voltage Fluctuations + 10%

Mains Transient Overvoltage Installation Category II (Per IEC 664)

Pollution Degree 2 (Normally only dry pollution, but with temporary conductivity caused by condensation)

NOTE:Tests have shown that because of the sensitive nature of the measurement and logic circuits, the following precautions must be observed:

Minimize exposure to ESD Events. If the M.O.L.E. Profiler or one of the thermocouples receives an 8kV electrostatic discharge during the data collection sequence, the M.O.L.E. Profiler may switch itself “OFF”. To retrieve the data, simply upload the data that was recorded prior to the event. If a new data collection sequence is started without uploading first, the existing data stored in the M.O.L.E. Profiler will be lost.

Keep strong electromagnetic fields away. The thermocouple wires serve as an antenna for electromagnetic radiation. If field strength of 3 volts per meter is present (usually due to close proximity of radio transmitters) while the M.O.L.E. Profiler is collecting data, the accuracy of the data may be compromised. This interference is compounded by using long thermocouple wires, as well as the effect of the thermocouple wires acting as a “tuned” antenna.