Wireless RF Option Setup

ECD Homepage  Help Index  Previous  Next

Navigation: Basics > Setup > Wireless RF Option Setup


To connect the wireless RF option:

Symbol_(Info)

Prior to  wireless RF communication, the M.O.L.E. Profiler must first be connected to a computer through the USB port. This allows the software to register the M.O.L.E. Profiler ensuring that other local computers cannot read RF data from this profiler. (Refer to Communications Setup for more information)

1)Insert the Wirless RF Transceiver into an available USB port and the antenna to the M.O.L.E. Profiler. The first time the Transceiver is connected to a computer, the drivers will automatically be installed. The operating system displays the installation status on the taskbar.

transceiver_install-1transceiver_install-2transceiver_install-3

mm_rf-pc_connection

Symbol_(Warning)

The M.O.L.E. Profiler is designed to work with with included antenna ECD P/N: E47-6342-11 only. Never make any changes or modifications to the Wireless RF components. Such changes are not expressly approved by ECD and could void the user's authority to operate the equipment.

2)After the Transceiver is connected to a computer, it needs to be detected by the software to properly communicate.
3)Start the software program by either double-clicking the MAP software icon or selecting it from the ECD program sub-menu.

63420005

4)On the M.O.L.E. menu, click the Select Instrument command.
5)Select the desired instrument from the list box. If there are none listed, click the Scan for Instruments command button to detect all available instruments.

Symbol_(Info)

When the software detects the Wireless RF Transceiver for the first time, the user will be prompted to enter the 5 digit Unlock Key located on the product ID label.

transeiver-back

6)Enter the Unlock Key in the field and select the OK command button.

63420005ab

Symbol_(Info)

MAP software allows multiple instruments to be connected to a computer at one time. Selecting the Scan for Instruments command button will detect all instruments and display them in the list. If no instrument is detected, the default Demonstration MEGAM.O.L.E.® profiler is displayed.

634200104a

7)Click the OK command button to accept or Cancel to quit the command.

The system is now ready for Wireless RF communication.