Changing the Calendar/Clock Battery

<< Click to Display Table of Contents >>

Navigation:  Instruments > SuperM.O.L.E.® Gold > Service and Troubleshooting >

Changing the Calendar/Clock Battery

If the clock is resetting itself, it may be necessary to replace the calendar/clock battery. (Note that changing this battery resets the hardware configuration parameters stored in the M.O.L.E. Profiler)

 

Change the battery as follows:

1)Remove the Power Pack by separating the unit. This exposes the clock battery.

smg_clock_battery

2)Pull the tab to remove the old battery.
3)Insert a new 3V Lithium battery (ECD P/N F30-0041-00), making sure it is oriented as pictured on the M.O.L.E. Profiler and making sure the tab is tucked in behind the battery.
4)To help prolong clock battery life, the Power Pack should be replaced so power is not drawing off the Calendar/Clock battery. The Activity LED blinks rapidly for two seconds when the Power Pack is first reconnected indicating complete reset has occurred.
5)Start and then stop the M.O.L.E. Profiler to clear the memory.
6)Reconfigure the M.O.L.E. Profiler Refer to Basics>Setup>Communications Setup for more information).